Principle
Platelet-poor plasma is added to an equal volume of partial thromboplastin reagent and warmed to 37°C for an exact incubation time. Pre-warmed (37°C) calcium chloride reagent (0.025M) is added to this mixture to activate the coagulation cascade. The time required for clot formation is recorded. Clot formation may be detected by optical or electromechanical methods using manual, semi-automated, or automated devices.
Reagents and Equipment
- Partial thromboplastin reagent: consists of phospholipids and a contact activator
- Calcium chloride reagent, 0.025M
- Fibrometer system
a. Fibrometer
b. Thermal prep block, 37 + 1°C
c. Automatic pipet - Coagulation cups, non-wettable surface
- Fibro-tips
Quality control materials (normal and abnormal) with established control limits should be run. NCCLS recommends controls be tested at the beginning of each testing day, followed by testing during each subsequent shift or with each batch of assays.
Specimen
Whole blood anticoagulated with 3.2% sodium citrate is the specimen of choice. The specimen should be processed to obtain platelet-poor plasma.
Procedure (Fibrometer Testing)
- Pre-warm a sufficient quantity of calcium chloride reagent to 37°C for the number of tests to be performed.
- Pipet 0.1 mL patient sample or control sample into labeled coagulation cup using the automatic pipet.
- To each sample cup, add 0.1 mL of partial thromboplastin
reagent using the automatic pipet. Carefully mix the contents of each
cup.
- Allow sample-partial thromboplastin reagent mixture to pre-warm to 37°C for 1-2 minutes. No longer than five minutes.
- Pipet 0.1 mL pre-warmed calcium chloride reagent into
coagulation cup containing sample-partial thromboplastin reagent mixture
using the automatic pipet (switch in the "on" position). When the
calcium chloride reagent is dispensed the timer will automatically
start. Alternatively, the timer may be started by touching the timer
plate. Reagent should be forcibly added to ensure mixing of reagent and
sample.
- The timer will stop when clot formation occurs.
- Record time taken for clot formation.
- Each sample (patient or control) should be run in duplicate.
The average of the duplicate partial thromboplastin times are recorded to the nearest second.
Reference Interval
Each laboratory should establish its own reference interval following a recommended procedure. The reference interval should be re-established with changes in instrumentation, reagent lot number, or at least once a year.
Comments
1- Partial thromboplastin reagent consists of phospholipids and a
contact activator. Kaolin, celite, silica, and ellagic acid are examples
of activators available. Care should be taken in choosing the partial
thromboplastin reagent, since reagents vary in their sensitivity or
insensitivity to lupus anticoagulant.
2- Each laboratory should determine the optimal incubation time
for partial thromboplastin reagent and sample based on its assay
procedure. The longer the incubation time, the shorter the partial
thromboplastin times due to increased contact activation. Excessive
heating (>5 minutes) will result in loss of Factor V and Factor VIII.
3- Precision between duplicate measurements is said to be
acceptable if the difference between duplicates is 10% or less of the
mean of the duplicates.
4- The APTT has been used to monitor heparin therapy. However,
newer methodologies such as Factor Xa inhibition assay are replacing
this use of the APTT.
5- The APTT is prolonged in:
C = (1.85 x 10-3) X (100 - Hct) X V
Where: C = volume of sodium citrate, V = volume of whole blood drawn, Hct = patient's hematocrit.
- Inherited single factor deficiencies of factors XI, X, IX, VIII, V, II, and I.
- Disseminated intravascular coagulation (DIC).
- Presence of circulating inhibitors like lupus-like anticoagulant.
C = (1.85 x 10-3) X (100 - Hct) X V
Where: C = volume of sodium citrate, V = volume of whole blood drawn, Hct = patient's hematocrit.
Note
Potential sources of error
- Associated with specimen
inappropriate ratio of anticoagulant to blood
failure to correct citrate volume if hematocrit >55%
clotted, hemolyzed, icteric, or lipemic specimen
delay in processing or testing
inappropriate storage
- Associated with reagents
incorrect preparation of reagents
failure to properly store reagents
use of reagents beyond reconstituted stability time
use of reagents beyond expiration date
contaminated reagents - Associated with procedureincorrect temperatureincorrect incubation timesincorrect volumes of sample, reagents, or bothimproperly functioning instrument
0 comments:
Post a Comment